DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

F2813-22 查看數據表(PDF) - Hamamatsu Photonics

零件编号
产品描述 (功能)
生产厂家
F2813-22
Hamamatsu
Hamamatsu Photonics Hamamatsu
F2813-22 Datasheet PDF : 6 Pages
1 2 3 4 5 6
OPERATING PRINCIPLE
As shown in the figure, when a voltage VD is supplied across the
input-side and output-side electrodes of the MCP, a potential
gradient is built up along the channel direction. If an incident
electron strikes an inner wall on the input side, a number of
secondary electrons are emitted. These secondary electrons
are accelerated by the potential gradient and travel along a
parabolic path determined by the initial velocity. They then
collide with the opposing wall surface, causing secondary
electrons to be emitted again. In this manner, the electrons
collide repeatedly within the channel as they pass towards the
output side. The result is a logarithmic multiplication of the
electron current appearing at the output.
MCP Configuration
PRIMARY
ELECTRON
CHANNEL WALL
STRIP CURRENT
OUTPUT
ELECTRON
VD
CURRENT GAIN AND LIFE CHARACTERISTICS
Current Gain vs. Supply Voltage
Life
TMCPB0029EA
105
TMCPB0030EC
1.0
TMCPC0030EA
104
0.5
103
SUPPLY VOLTAGE : 1000 V
OUTPUT CURRENT DENSITY : 50 nA/ 15 mm
102
600
800
1000
1200
SUPPLY VOLTAGE (V)
0
10–4
10–3
10–2
10–1
ACCUMULATED CHARGE OF OUTPUT SIGNAL (C/cm2)
s PRECAUTIONS FOR USE
Don’t touch the MCP and assembly parts with bare hands.
Always handle the MCP in an area that is maintained at the same level of cleanliness as a clean room since dust and
humidity may adversely affect MCP characteristics.
The MCP should be kept in vacuum or nitrogen atmosphere when storing it after delivery.
If excessive outgas from the MCP occurs, perform vacuum-baking of the MCP at +350°C maximum in an evacuating
system. Alternatively, electron bombarding of the MCP is effective. (Please consult our sales office for details.)
The MCP and MCP assemblies must be operated in a vacuum below 1.3 × 10-3 Pa (1 × 10-5 Torr). However, phosphor
output MCP assemblies and other assemblies using a three-stage MCP should be operated in a vacuum below
1.3 × 10-4 Pa (1 × 10-6 Torr).

Share Link: 

datasheetq.com  [ Privacy Policy ]Request Datasheet ] [ Contact Us ]