DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

PBM25US28(Rev3_1) 查看數據表(PDF) - XP Power Limited

零件编号
产品描述 (功能)
生产厂家
PBM25US28
(Rev.:Rev3_1)
XPPOWER
XP Power Limited XPPOWER
PBM25US28 Datasheet PDF : 2 Pages
1 2
AC-DC Medical
25 Watts
PBM25 Series
THE XPERTS IN POWER
Small Size 2.50” x 4.25” x 0.96”
Low Leakage Current
Medical Safety Agency Approvals
Both Line and Neutral Fused
Meets Class B, EN55011 Conducted
Fixed Switching Frequency
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Earth
Leakage Current
Output
Output Voltage
Output Power
Minimum Load
Start Up Delay
Start Up Rise TIme •
Hold up Time
Initial Set Accuracy •
Line Regulation
Total Regulation
Transient Response •
Ripple & Noise
Overvoltage
Protection
Overcurrent
Protection
Short Circuit
Protection
Temperature
Coefficient
85-264 VAC
47-63 Hz
0.7 A (rms) at 115 VAC,
0.4 A (rms) at 230 VAC
10 A max at 115 VAC,
25 A max at 230 VAC
90 µA max at 115 VAC 60 Hz
150 µA max at 230 VAC 50 Hz
5.1 to 28 VDC, See Table
25/28 Watts, See Table
No minimum for single outputs models
See Table for multi outputs models
3 s max
12 ms
10 ms minimum at 110 VAC
±1%
±0.5% max at full load
See Table
4% max deviation, 500 µs recovery
time for a 25% load change
±1% max pk-pk (See Note 1)
Provided on output 1 only; set at
112-132% of its nominal output
voltage
125-200% with auto recovery
Trip & restart with auto recovery
±0.04%/°C
General
Efficiency
Isolation
Switching Frequency •
MTBF
Size
Weight
68% minimum for 5 V & multi output
models, 70% minimum all others
4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
42 KHz ±5 KHz
500,000 hrs per min MIL-HDBK-217F
4.25 x 2.50 x 0.96”
107.95 x 63.50 x 24.38 mm
Single output models 163 g,
Multi output models 175 g
Environmental
Operating
Temperature
Storage Temperature •
Cooling
Relative Humidity •
Operating Altitude •
Shock
Vibration
0 °C to +70 °C,
derate linearly from 100% load at
+50°C to 50% load at +70 °C
-40 °C to +85 °C
Convection cooled
5-95%, non-condensing
3000m
20 G in 3 directions, 3 times each
0.15mm over 10-55Hz and 1
octave/min. 3 axes for 30 min
EMC & Safety
EMC
ESD
Radiated Immunity •
EFT/Burst
Surge
Safety Approvals
EN55011 (CISPR11) Class B & FCC
Class B conducted
EN61000-4-2 Level 2 Perf Criteria A
EN61000-4-3 3 V/m Perf Criteria A
EN61000-4-4 Level 3 Perf Criteria A
EN61000-4-5 Level 3 Perf Criteria A
UL2601-1, CSA C22.2 No 601.1,
EN60601-1, IEC 601-1
www.xp-medical.com
sales@xp-medical.com