PA4000 Power Analyzer
Effect of Common Mode
(Typical)
1000 V, 60Hz < 10 mV
100 V, 100 kHz < 50 mV
Current Accuracy, Arms ( 45 Hz ± 0.04% of Reading ± 0.04% of Range ± (20 µV/Zext)
- 850 Hz )
Current Accuracy, Arms ( 10 Hz ± 0.05% of Reading ± 0.05% of Range ± (0.02*F)% of Reading ± (20 µV/Zext) (typical)
- 45 Hz, 850 Hz - 1 MHz )
Current Accuracy, DC
Current Accuracy, Armn
± 0.05% of Reading ± 0.1% of Range ± (50 µV/Zext)
± 0.2% of Reading ± 0.1% of Range ± (100 µV/Zext)
Measurement accuracy - power
Watts Accuracy
VA Accuracy
VAR Accuracy
(Vrms acc. x Arms x PF) ± (Arms acc. x Vrms x PF) ± (Vrms x Arms x cos θ -cos { θ± (Vh1ph.err. ± Ah1ph.err.)})
(Vrms acc. x Arms) ± (Arms acc. x Vrms )
PF Accuracy
Cos θ -cos [ θ± ( Vh1 ph.err± Ah1 ph.err)] ± 0.001
Measurement accuracy - Harmonic
magnitude and phase
Voltage Harmonics (45 Hz -
850 Hz) (Typical for 10 Hz to
45 Hz and 850 Hz to 1 MHz)
± 0.08% of Reading ± 0.08% of Range ± (0.02*F)% of Reading ± 0.02 V
Voltage Harmonics Phase
(45 Hz - 850 Hz) (Typical for
10 Hz to 45 Hz and 850 Hz to
1 MHz)
0.025 ± [ 0.005 * Vrange / Vreading )] ± (0.1 / Vrange) ± ( 0.002 *F )
Current Harmonics (45 Hz -
850 Hz) (Typical for 10 Hz to
45 Hz and 850 Hz to 1 MHz)
± 0.08% of Reading ± 0.08% of Range ± (0.02*F)% of Reading ± (20 µV /Zext)
Current Harmonics Phase
(45 Hz - 850 Hz) (Typical for
10 Hz to 45 Hz and 850 Hz to
1 MHz)
0.025 ± [0.005 * Arange / Areading)] ± (0.001 / Arange * Zext)) ± (0.002 *F)
Physical characteristics
Dimensions
Weight
Temperature
Height
Width
Depth
Net (without lead set)
Storage
Operating
13.2 cm.
42 cm.
31 cm.
8.8 kg
5.2 in.
16.5 in.
12.5 in.
19.5 lb.
-20 ○ C to +60 ○ C
0 ○ C to +40 ○ C
www.tektronix.com 5