DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

PMP110-42 查看數據表(PDF) - Tumbler Technologies + TRUMPower

零件编号
产品描述 (功能)
生产厂家
PMP110-42
TRUMPOWER
Tumbler Technologies + TRUMPower TRUMPOWER
PMP110-42 Datasheet PDF : 2 Pages
1 2
110 WATT MEDICAL SWITCHING POWER
DESCRIPTION
The PMP110 series of AC/DC switching power supplies are for
110 watts of continuous output power. They are enclosed in a
94V-1 rated polyphenylene-oxide case with an IEC 320 inlet to
mate with interchangeable cord for world-wise use. An on/off
switch can be added at request. All models meet CISPR 11 and
FCC class B emission limits, and are designed for medical
applications not for patient contact.
PMP110 SERIES
(MDD+EMC)
FEATURES
Dual, triple or quadruple outputs
Optional output connectors
Optional on/off switch
Optional Power Fail Detect (PFD) signal
100% burn-in
W i d e i n p u t r a n g e 8 5 t o 2 6 4 VA C
Input surge current protection
Overvoltage protection
Overcurrent protection
INPUT SPECIFICATIONS
Input voltage
8 5 t o 2 6 4 VA C
Input frequency47 to 63 Hz
Input current
3 . 2 0 A ( r m s ) f o r 11 5 VA C
1 . 8 0 A ( r m s ) f o r 2 3 0 VA C
L e a k a g e c u r r e n t 9 0 u A m a x . @ 11 5 VA C , 6 0 H z
1 5 0 u A m a x . @ 2 3 0 VA C , 5 0 H z
Safety Standard Approvals
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
ENVIRONMENTAL SPECIFICATIONS
Operating temperature0to +70
Storage temperature
-40to +85
Relative humidity
5to 95non-condensing
Derating
Derate from 100at +50
linearly to 50at +70
OUTPUT SPECIFICATIONS
Output voltage/current
Total output power
Ripple and noise
Overvoltage protection
See rating chart
110 watts maximum
1peak to peak maximum
Provided on output #1 only; set
at 112 - 132 of its nominal
output voltage
Overcurrent protection
All outputs protected to short
circuit conditions
Temperature coefficientAll outputs ±0.04 /
maximum
Transient response
Maximum excursion of 4 or
better on all models, recovering
to 1 of final value within
500us after a 25step load
change
PFD signal (optional):
TTL logic high for normal
operation and TTL logic low
upon loss of input power. This
signal appears at least 1ms prior
to +5V output dropping 5%
below its nominal value. This
signal also provides a minimum
delay of 100ms after +5V is
within regulation.
GENERAL SPECIFICATIONS
Switching frequency20KHz to 250KHz, varies with load and
line
Efficiency
70minimum on single output models
with Vo 12V, 65% minimum on the
others
Hold-up time
12 msec minimum at 110VAC
Line regulation
±0.5maximum at full load
Inrush current
15 amps @ 115VAC, or 30 amps @
230VAC, at 25cold start
Withstand voltage
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF
550,000 hours minimum at full load at
25ambient , calculated per
MIL-HDBK-217F, excluding DC fan.
EMC Performance(EN60601-1-2: 2001)
EN55011:
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Harmonic distortion , Class A
Line flicker
ESD,± 8KV air and ± 4KV contact
EN61000-4-3:
Radiated immunity,3V/m for 80~2500MHz
EN61000-4-4:
Fast transient/burst, ± 1KV
EN61000-4-5:
Surge, ± 1KV diff., ± 2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
EN61000-4-11:
Magnetic field immunity, 1A/m
Voltage dips, 30reduction for 500ms and
>95 reduction for 10ms Interruptions,
>95for 5000ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 sales@trumpower.com www.TRUMPower.com

Share Link: 